Page 1 - MDP 5011

MDP 5011Molecular PumpUser’s Manual

Page 2

A 2102 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualThe different version operatingprincipleMPD 5011CP VERSIONFOR CORROSIVE GASESThe MD

Page 3 - Molecular pumps

A 4001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 technicalcharacteristicsCharacteristics Units MPD 5011 MDP 5011CPInlet fla

Page 4

A 4011/3Alcatel Vacuum Technology France - MDP 5011 User’s ManualPump dimensionsEdition 01 - Sep 2004Gb 02560Power supply - ConnectorDIMENSIONSMDP 501

Page 5 - Contents

A 4012/3Alcatel Vacuum Technology France - MDP 5011 User’s ManualPump dimensionsEdition 01 - Sep 2004GB 02560DIMENSIONSMDP 5011 DN 63 ISO-Kwater coole

Page 6

A 4013/3Alcatel Vacuum Technology France - MDP 5011 User’s ManualPump dimensionsEdition 01 - - Sep 2004GB 02560DIMENSIONSMDP 5011 DN 63 ISO-Kair coole

Page 7 - Chapter A

A 5101/1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 accessoriesSCREEN FILTER: 2,5 MMThis filter protects the pump against solid

Page 8 - Edition 02 - March 06GB 02618

Chapter B1 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manual ofMDP 5011 molecular pumpsSTART-UPSafety instructions B 100. . .

Page 9 - Edition 02 - March 06GB 02596

B 1001 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualSafety instructionsIndicates a potentially hazardous situation which, if notavoided

Page 10 - GB 02596

B 1002 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualSafety instructionsINSTALLATION - START UPThe controllers are designed to guarantee

Page 11 - Edition 02 - March 06GB 02599

B 1003 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualSafety instructionsINSTALLATION - START UP(CTD)The vacuum pump is also a compressor

Page 12 - Pump dimensions

Alcatel Vacuum Technology, as part of the AlcatelGroup, has been supplying vacuum pumps, leakdetection systems, vacuum measurement andmicro machining

Page 13

B 3001 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualPump connection toan installationThe pump can operate in anypositionIn.In.In.In.The

Page 14

B 3002 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualPump connection toan installationCOOLING RECOMMENDATIONSAND TEMPERATURE LIMITSAmbie

Page 15 - MDP 5011 accessories

B 3201 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualNitrogen purge connection(MDP 5011CP)For proper protection from reactive gases and

Page 16 - Chapter B

B 3301/1Alcatel Vacuum Technology France - MDP 5011 User’s ManualVenting valve connectionCheck the solenoid valve voltage : it must be compatible with

Page 17 - Safety instructions

B 3401 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualWater cooling connectionCHARACTERISTICS OF WATERCOOLINGIn order to limit the corros

Page 18 - Edition 03 - March 06

B 3402 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualWater cooling connectionConnection to the water line- Use a flexible or a rigid 6 x

Page 19 - Edition 03 - March 06GB 00725

B 3501 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualAir cooling kit connectionAir cooling kit:When the water cooling kit must be replac

Page 20 - Edition 01 - Sep 2004GB 02602

Chapter C1 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manual ofMDP 5011 molecular pumpsOPERATIONSafety instructions for produc

Page 21

C 1001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualSafety instructions for productuseBefore to use the controller, make sure that the

Page 22 - Edition 02 - March 06GB 03078

C 1501 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMolecular pump operation in apumping applicationFIRST PUMP START-UPWhen the pump is

Page 23 - Venting valve system

1/ 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011Molecular pumpsWELCOMEDear customer,You have just purchased an AdixenMolecular pu

Page 24 - Water cooling connection

C 1502 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualMolecular pump operation in apumping applicationEXAMPLE OF A 3 VALVEASSEMBLY (EXAMP

Page 25 - GB 002603

Chapter D1 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manual ofMDP 5011 molecular pumpsMAINTENANCEMDP 5011pump maintenance fre

Page 26 - Edition 02 - March 06Gb 02605

D 3001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 pump maintenancefrequencyHousing temperature °CHousing temperature measuri

Page 27 - Edition 02 - march 06GB 02626

E 1001/1Alcatel Vacuum Technology France - MDP 5011 User’s ManualSafety recommendationsrelated to m aintenanceStandard precautions before any maintena

Page 28

E 2001 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 pump lubricationThe first lubrication required for the correct operation o

Page 29 - Edition 02 - March 06Gb 02608

E 2002 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 pump lubrication36415812FKFRELUBRICATION PROCESSBearing oppositethe ”pumpi

Page 30 - Gb 02608

E 2003 / 3Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 pump lubricationBEARING LUBRICATION- Stop the MDP pump and the primary pum

Page 31 - Edition 02 - March 06GB 02626

2nd cycle1st cycleE 3001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMPD 5011 running-in usingACT 100 controllerStart the MDP to atmos

Page 32 - MDP 5011 pump maintenance

F 0001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualSpare parts - Instructions of useREPLACEMENT OF PARTS ANDUSE OF NON GENUINE PARTSOu

Page 33 - Safety recommendations

F 1001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 first level ofmaintenanceLUBRICATION SYRINGEReady-to-use, it contains the

Page 34 - Edition 02 - March 06GB 02611

2/ 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011Molecular pumpsThis product complies with requirements of EuropeanDirectives, lis

Page 35 - GB 02611

G 010Alcatel Vacuum Technology France - MDP 5011 User’s ManualPumping curvesPumping speed0123456781,00E--04 1,00E--03 1,00E--02 1,00E--01 1,00E+00 1,0

Page 36 - Edition 02 - March 06Gb 02611

G 0102/2Alcatel Vacuum Technology France - MDP 5011 User’s ManualPumping curvesFlow1,00E--021,00E--011,00E+001,00E+011,00E+021,00E+031,00E--04 1,00E--

Page 37 - ACT 100 controller

G 2001/2Alcatel Vacuum Technology France - ACT 100/ACT 100T User’s manualSafety questionnaireProcedure for returning ADIXEN vacuum pumpsand helium lea

Page 38 - Edition 02 - Sept 03GB 01390

G 2002/2Alcatel Vacuum Technology France - ACT 100/ACT 100T User’s manualSafety questionnaireCe questionnaire est téléchargeable sur le site : www.adi

Page 39 - Edition 02 - March 06Gb 02613

CHINAAlcatel Vacuum Technology ShangaiN°82 Lane 887 Zuchongzhi RoadZhangjiang High-Tech Park,Shanghai 201203 - P.R. ChinaTel. (86) 21 5027 0628Fax (86

Page 40 - MDP5011 with RVP 2m

Contents1/2Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manualMolecular pumpsINTRODUCTIONMDP 5011 molecular pump overview A 200...

Page 41 - Pumping curves

Contents2 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manualMolecular pumpsMAINTENANCE COMPONENTSSpare parts - Instruction of u

Page 42 - Safety questionnaire

Chapter A1 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualUser’s manual ofMDP 5011 molecular pumpsINTRODUCTIONMDP 5011 molecular pump ove

Page 43

A 2001 / 1Alcatel Vacuum Technology France - MDP 5011 User’s ManualMDP 5011 molecular pumpoverviewTWO PUMP VERSIONSThe standard version : MDP 5011 pum

Page 44

A 2101 / 2Alcatel Vacuum Technology France - MDP 5011 User’s ManualThe different version operatingprincipleMDP 5011 MOLECULAR PUMPThe rotor, a smooth

Comments to this Manuals

No comments